How To Calibration FOX-3D Alignment Machine

Image
  Details: Service Instruction as been released with the Single shaft calibration Procedure for FOX 3D WA Machines. Procedure has been given below for your easy understanding Assembly procedure: 1. Mount the Calibration Shaft with the Calibration kit stands at both ends. Ensure the horizontality of shaft using Spirit level by adjusting the Jack screws provided in the Stands and then lock the position using Lock nuts. Finally lock the Shaft with Stands firmly using M8x25 Allen screws to avoid rotation as shown below: Cover Cover both the ends of Calibration shaft using End cap. Calibration Calibration procedure: 1. Select the Single shaft calibration Icon in the Settings screen: ← 2. Following screen will be displayed. Follow the On-screen instructions 3. Place the calibration kit on the Rotary plates (turn tables) at a distance o...